:SYST:SEC:LEV
Syntax :SYSTem:SECurity[:LEVel] {NON|LOW|HIGH}
:SYSTem:SECurity[:LEVel]?
Description Sets/Reads the security level.
Parameters
When the setting of security level is LOW, it is able to change to NON or HIGH. But when
this setting is HIGH, it is not able to change NON or LOW.
The setting of security level can be turned NON by executing the preset or recalling when
the setting of security level is HIGH.
Even if the setting of security level is ON, the command that reads out the frequency is not
influenced.
Query response {NON|LOW|HIGH}<newline><^END>
Example of use
10 OUTPUT 717;":SYST:SEC LOW"
20 OUTPUT 717;":SYST:SEC?"
30 ENTER 717;A$
Equivalent key [System] - Service Menu - Security Level - None|Low|High
Description
NON (preset value) Turns OFF the security level.