User`s guide

Table Of Contents
Chapter 13 437
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:SOUR{1-4}:POW:SLOP:STAT
Syntax :SOURce{[1]|2|3|4}:POWer[:LEVel]:SLOPe:STATe {ON|OFF|1|0}
:SOURce{[1]|2|3|4}:POWer[:LEVel]:SLOPe:STATe?
Description Turns on/off the power slope feature of channel 1 (:SOUR1) to channel 4 (:SOUR4). This
function is a function to correct the attenuation of simple power level proportional to the
frequency (attenuation due to cables and so on).
Parameters
Query response {1|0}<newline><^END>
Example of use
10 OUTPUT 717;":SOUR1:POW:SLOP:STAT ON"
20 OUTPUT 717;":SOUR1:POW:SLOP:STAT?"
30 ENTER 717;A
Related commands :SOUR{1-4}:POW:SLOP on page 436
Equivalent key
[Sweep Setup] - Power - Slop [ON/OFF]
Description
ON or 1 Turns on the power slope feature.
OFF or 0 (preset value) Turns off the power slope feature.