E5061A/E5062A commands
13. SCPI Command
Reference
:SENS{1-4}:CORR:COLL:SAVE
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect:SAVE
Description From the measured calibration data, calculates the calibration coefficients depending on
the selected calibration type.
Calculating the calibration coefficients clears all calibration data whether or not used for
the calculation and also clears the calibration type selections.
If you execute this command before all necessary calibration data for calculating the
calibration coefficients is measured, an error occurs and the command is ignored. (No
query)
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:SAVE"
Related commands :SENS{1-4}:CORR:COLL:METH:OPEN on page 402
:SENS{1-4}:CORR:COLL:METH:SHOR on page 402
:SENS{1-4}:CORR:COLL:METH:THRU on page 404
:SENS{1-4}:CORR:COLL:METH:SOLT1 on page 403
:SENS{1-4}:CORR:COLL:METH:SOLT2 on page 403
Equivalent key
[Cal] - Calibrate - Response|n-Port Cal - Done
:SENS{1-4}:CORR:COLL:SHOR
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect[:ACQuire]:SHORt <numeric>
Description For channel 1 (:SENS1) to channel 4 (:SENS4), measures the calibration data of the short
standard of the specified port. (No query)
Parameters
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:SHOR 1"
20 OUTPUT 717;"*OPC?"
30 ENTER 717;A
Equivalent key [Cal] - Calibrate - Response (Short)|1-Port Cal - Short
[Cal] - Calibrate - n-Port Cal - Reflection - Port m Short
<numeric>
Description Port number
Range 1 to 2
Resolution 1