User`s guide

Table Of Contents
Chapter 13 393
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:SENS{1-4}:CORR:COLL:CKIT:STAN{1-21}:TYPE
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect:CKIT:STAN{[1]|2|3|4|5|6|7|8|9|10|11|12|13|14|1
5|16|17|18|19|20|21}:TYPE {OPEN|SHORt|LOAD|THRU|ARBI|NONE}
:SENSe{[1]|2|3|4}:CORRection:COLLect:CKIT:STAN{[1]|2|3|4|5|6|7|8|9|10|11|12|13|14|1
5|16|17|18|19|20|21}:TYPE?
Description For the calibration kit selected for channel 1 (:SENS1) to channel 4 (:SENS4), sets the
standard type of the standard1 (:STAN1) to standard 21 (:STAN21).
Parameters
Query response {OPEN|SHORt|LOAD|THRU|ARBI|NONE}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:CKIT:STAN1:TYPE OPEN"
20 OUTPUT 717;":SENS1:CORR:COLL:CKIT:STAN1:TYPE?"
30 ENTER 717;A$
Equivalent key [Cal] - Modify Cal Kit - Define STDs - no. name
*1
- STD Type
Description
OPEN Specifies open.
SHORt Specifies short.
LOAD Specifies load.
THRU Specifies thru.
ARBI Specify arbitrary impedance.
NONE Specifies DUT of which theoretical values are 0.
*1.no: standard number (1 to 21), name: standard name (variable)