User`s guide

Table Of Contents
Chapter 13 391
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:SENS{1-4}:CORR:COLL:CKIT:STAN{1-21}:LAB
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect:CKIT:STAN{[1]|2|3|4|5|6|7|8|9|10|11|12|13|14|1
5|16|17|18|19|20|21}:LABel <string>
:SENSe{[1]|2|3|4}:CORRection:COLLect:CKIT:STAN{[1]|2|3|4|5|6|7|8|9|10|11|12|13|14|1
5|16|17|18|19|20|21}:LABel?
Description For the calibration kit selected for channel 1 (:SENS1) to channel 4 (:SENS4), sets the
name of the standard1 (:STAN1) to standard 21 (:STAN21).
Parameters
Query response {string}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:CKIT:STAN1:LAB ""OPEN"""
20 OUTPUT 717;":SENS1:CORR:COLL:CKIT:STAN1:LAB?"
30 ENTER 717;A$
Equivalent key [Cal] - Modify Cal Kit - Define STDs - no. name
*1
- Label
<string>
Description Standard name
Range 254 characters or less
Preset value Varies depending on the calibration kit and the standard.