User`s guide

Table Of Contents
Chapter 13 381
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:SENS{1-4}:CORR:COLL:CKIT:STAN{1-21}:ARB
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect:CKIT:STAN{[1]|2|3|4|5|6|7|8|9|10|11|12|13|14|1
5|16|17|18|19|20|21}:ARBitrary <numeric>
:SENSe{[1]|2|3|4}:CORRection:COLLect:CKIT:STAN{[1]|2|3|4|5|6|7|8|9|10|11|12|13|14|1
5|16|17|18|19|20|21}:ARBitrary?
Description For the calibration kit selected for channel 1 (:SENS1) to channel 4 (:SENS4), sets the
value of the arbitrary impedance of the standard1 (:STAN1) to standard 21 (:STAN21).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:CKIT:STAN1:ARB 50.5"
20 OUTPUT 717;":SENS1:CORR:COLL:CKIT:STAN1:ARB?"
30 ENTER 717;A
Equivalent key [Cal] - Modify Cal Kit - Define STDs - no. name
*1
- Arb. Impedance
<numeric>
Description Arbitrary impedance
Range -1E18 to 1E18
Preset value Varies depending on the calibration kit and the standard.
Unit Ω (ohm)
*1.no: standard number (1 to 21), name: standard name (variable)