User`s guide

Table Of Contents
376 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:SENS{1-4}:CORR:COLL:CKIT:LAB
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect:CKIT:LABel <string>
:SENSe{[1]|2|3|4}:CORRection:COLLect:CKIT:LABel?
Description Sets a calibration kit name for the calibration kit selected for channel 1 (:SENS1) to
channel 4 (:SENS4).
Parameters
Query response {string}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:CKIT:LAB ""MY_KIT"""
20 OUTPUT 717;":SENS1:CORR:COLL:CKIT:LAB?"
30 ENTER 717;A$
Related commands :SENS{1-4}:CORR:COLL:CKIT on page 375
Equivalent key
[Cal] - Modify Cal Kit - Label Kit
<string>
Description Calibration kit name
Range 254 characters or less
Preset value Varies depending on the calibration kit number as follows:
1: “85033E”
2: “85033D”
3: “85052D”
4: “85032F”
5: “85032B”
6: “85036B/E”
7: “85039B”
8: “85038A/F/M”
9 to 10: “User”