User`s guide

Table Of Contents
294 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:CALC{1-4}:SMO:APER
Syntax :CALCulate{[1]|2|3|4}[:SELected]:SMOothing:APERture <numeric>
:CALCulate{[1]|2|3|4}[:SELected]:SMOothing:APERture?
Description Sets the smoothing aperture for channel 1 (:CALC1) to channel 4 (:CALC4) (specified
with the :CALC{1-4}:PAR{1-4}:SEL command).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:SMO:APER 2.5"
20 OUTPUT 717;":CALC1:SMO:APER?"
30 ENTER 717;A
Related commands :CALC{1-4}:PAR{1-4}:SEL on page 286
:CALC{1-4}:SMO on page 293
Equivalent key
[Avg] - Smo Aperture
<numeric>
Description Percentage relative to the sweep span value
Range 0.05 to 25
Preset value 1.5
Unit % (percent)
Resolution 1E-14