User`s guide

Table Of Contents
Chapter 13 225
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:CALC{1-4}:FUNC:DOM:STOP
Syntax :CALCulate{[1]|2|3|4}[:SELected]:FUNCtion:DOMain:STOP <numeric>
:CALCulate{[1]|2|3|4}[:SELected]:FUNCtion:DOMain:STOP?
Description For channel 1 (:CALC1) to channel 4 (:CALC4), sets the stop value of the analysis range
of the :CALC{1-4}:FUNC:EXEC command.
When the trace coupling is off, the active trace is the target to be set.
Parameters
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:FUNC:DOM:STOP 1.8E9"
20 OUTPUT 717;":CALC1:FUNC:DOM:STOP?"
30 ENTER 717;A
Related commands :CALC{1-4}:FUNC:EXEC on page 225
:CALC{1-4}:FUNC:DOM on page 222
:CALC{1-4}:FUNC:DOM:COUP on page 223
:CALC{1-4}:FUNC:DOM:STAR on page 224
Equivalent key No equivalent key is available on the front panel.
:CALC{1-4}:FUNC:EXEC
Syntax :CALCulate{[1]|2|3|4}[:SELected]:FUNCtion:EXECute
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4) (specified with the
:CALC{1-4}:PAR{1-4}:SEL command), executes the analysis specified with the
:CALC{1-4}:FUNC:TYPE command. (No query)
Example of use
10 OUTPUT 717;":CALC1:FUNC:EXEC"
Related commands :CALC{1-4}:FUNC:TYPE on page 231
:CALC{1-4}:FUNC:DOM on page 222
:CALC{1-4}:PAR{1-4}:SEL on page 286
Equivalent key No equivalent key is available on the front panel.
<numeric>
Description Stop value of the analysis range
Preset value 0
Unit Hz (hertz), dBm or second