AGILENT E5071C ENA NETWORK ANALYZER
Flow Diagram of Procedure The operational flow of the software is depicted by the flowchart shown below.
Basic Calibration Before executing calibration, you need to select a calibration kit. If you use a calibration kit other than a predefined one, you need to define it. If the connector type of the standard of the calibration kit you use has polarity (the distinction between male and female), you need to change the standard class definition of the calibration kit depending on the standard you actually use. For more information, see Modifying Calibration Kit Definition.
makes it possible to perform measurements with the highest possible accuracy. A total of twelve error terms, six each in the forward direction and the reverse direction, are used in a calibration. 1. Press Cal key. 2. Click Cal Kit, then select the calibration kit. 3. If the name (label) of the calibration kit has been changed, the label is displayed as the soft key. 4.
Safety Safety Summary The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific WARNINGS elsewhere in this manual may impair the protection provided by the equipment. Such noncompliance would also violate safety standards of design, manufacture, and intended use of the instrument.
Direct current. On (Supply). Off (Supply). In-position of push-button switch. Out-position of push-button switch. A chassis terminal; a connection to the instrument's chassis, which includes all exposed metal structure. Stand-by.