Installation guide
Chapter 2 15
Installation
Environmental Requirements
2. Installation
Ventilation Requirements
To ensure safety requirements, the specifications and measurement accuracy of the
analyzer, you must keep environmental temperature within the specified range by
providing appropriate cooling clearance around the analyzer or, for the rackmount type, by
forcefully air-cooling inside the rack housing. For more information on environmental
temperature to satisfy the specifications and measurement accuracy of the analyzer, see the
specification in the E5071C Online Help.
When the environmental temperature around the analyzer is kept within the temperature
range of the operating environment specification (See Section
“Operating Environments”
on page 14), the analyzer conforms to the requirements of the safety standard.
Furthermore, under that temperature requirement, the analyzer still conforms to the
requirements of the safety standard even when placing the analyzer with cooling clearance
as follows:
Requirements
Back ≥ 180 mm
Sides ≥ 60 mm (both right and left)
Figure 2-1 Ventilation space at the installation
e5071cqe1001
Softkey On/Off
Capture
Softkey On/Off
Capture
+ 26 dBm RF 35 VDC Max
Avoid Static Discharge CAT
Σ
ACTIVE CH/TRACE
RESPONSE NAVIGATION
ENTRY
INSTR STATEMKR/ANALYSIS
STIMULUS
PORT 1 PORT 2 PORT 3
PORT 4
USB
PROBE POWER
OO OO
OO