Installation guide

Operation Verification
Power Level Accuracy Check
Chapter 328
Calculate the Actual Power Level
1. Calculate and record the Actual Power Level, in Table 3-3, as the sum of the Power Meter
Reading for
15 dBm and the Spectrum Analyzer Marker (dB) value.
For example:
Power Meter Reading for
15 dBm at 2.516 MHz =
14.95 dBm
Spectrum Analyzer Marker (dB) at 2.516 MHz and Power Level Setting at
85 dBm =
70.17 dB
Actual Power Level at 2.516 MHz and
85 dBm: (
14.95) + (
70.17) =
85.12 dBm
2. Compare the calculated Actual Power Level value to the corresponding limits.