Operation Verification
Power Level Accuracy Check
Chapter 3 25
Power Level Accuracy Check
Performing this check will provide a high level of confidence that the signal generator’s power
level circuitry is functioning correctly. This check does not test the signal generator to
warranted specifications. Test points have been reduced and the limits are degraded in order
to minimize measurement time and take into account a broad range of measurement
uncertainties.
In order to have the signal generator tested to warranted specifications, a complete power
level accuracy performance test is required. If the complete performance test is needed and
you are unable to perform it, contact your nearest Agilent Technologies service center for
information concerning calibration. Refer to “Contacting Agilent Technologies” on page 17 for
contact information.
Connect the Test Equipment for Setup 1
Figure 3-2 Power Level Accuracy Equipment Setup 1
Configure the Signal Generator
1. Press Preset.
2. Press
Frequency. Enter 277 using the numeric keypad and press the kHz terminator
softkey.
3. Press
Amplitude. Enter 13 using the numeric keypad and press the dBm terminator softkey.
4. Press
Mod On/Off. The MOD OFF annunciator is displayed.
5. Press
RF On/Off. The RF ON annunciator is displayed.