Installation guide
Operation Verification
Operation Verification
Chapter 320
Operation Verification
Operation verification is a series of tests that, when completed, will either ensure that the
signal generator is operating correctly, or will assist in pointing to the problem area.
Operation verification does not ensure performance to specifications, but should provide a
level of confidence that the signal generator is operating correctly within a minimum amount
of time.
Operation verification is appropriate for incoming inspection, after repair (when a full
calibrated performance is not required), or whenever the integrity of the signal generator is in
question.
Perform the following tests in the order they are presented. The tables referenced by the tests
are located in the rear of the chapter, where they can be copied easily.
1. Power On the Signal Generator on page 21
2. Check for Error Messages on page 22
3. Frequency Range and Accuracy Check on page 23
4. Power Level Accuracy Check on page 25
5. FM Accuracy Check on page 29
6. AM Accuracy Check on page 31
7. I/Q Modulation Check on page 33
Required Equipment
• Agilent 53132A Option 050 Frequency Counter
• Agilent E4418B or E4419B Power Meter
• Agilent E9304A Power Sensor
• Agilent 8563E Spectrum Analyzer
• Agilent 8491A/B Option 006 Attenuator (6 dB)
• Agilent 8491A/B Option 010 Attenuator (10 dB)
• Agilent 8902A Measuring Receiver