Installation guide

Troubleshooting
Troubleshooting Unlevels and the RF Path
1-74
Troubleshooting a Leveling Problem
NOTE Before proceeding to troubleshoot with this section, run a full self-test and correct any reported
failures and unlock problems. (Refer to “Self-Test Failures and Related Troubleshooting” on
page 1-19 and “Troubleshooting Unlocks” on page 1-68.)
Check the RF path power level against the values in Table 1-33.
If the instrument cannot achieve the maximum power levels listed in Table 1-33,
perform the ALC Calibration and the Power Level Accuracy, High Power Calibration.
(Refer to the Calibration Guide.)
If these adjustments fail or the instrument still cannot achieve the maximum power levels listed in
Table 1-33, proceed to “Troubleshooting the RF Path” on page 1-76.
Check for an Unleveled Annunciator
If the Unleveled Annunciator is on:
1. Check that the signal generator's amplitude is not set higher than the maximum level specified on
the Specifications data sheet.
If the unleveled indication turns off after resetting the amplitude, the signal generator is
operating correctly.
2. Check that the signal generator's RF output is terminated into 50 ohms.
If the unleveled indication turns off after terminating the RF output into 50 ohms, the signal
generator is operating correctly.
If the Unleveled Annunciator remains on:
1. Use a spectrum analyzer to check the RF signal level at the RF Output connector.
Signal Generator Setup
a. Set
ALC On/Off to Off
b. Set
Amplitude to +25 dBm
c. Set
Modulation to Off
d. Set
RF On/Off to On
e. Set
Sweep to Frequency
f. Set
Sweep Type to Step
Table 1-33
Frequencies Standard Power Option UNB
250 kHz to 1 GHz +13 dBm +17 dBm
> 1 GHz to 3 GHz +10 dBm +16 dBm
> 3 GHz to 4 GHz +7 dBm +13 dBm
> 4 GHz to 6 GHz (Option 506) N/A +10 dBm