Installation guide

Troubleshooting
Getting Started with Troubleshooting
1-11
Check that there is approximately 290 Vac, 40 kHz signal at CN2-1. (This check should be made with the
A4 Inverter connected to the A2 Display.)
If the signal is present at the A4 Inverter output, replace the A2 Display.
(Refer to A2 Display” on page 2-10.)
If the signal is not present at CN2, use a DVM to measure the signals shown in Table 1-4 at CN1 on the
A4 Inverter. Probe CN1 from the top of the connector while it is still connected to the A3 Front Panel
Auxilary Board.
If the signals in Table 1-4 are correct, replace the A4 Inverter.
(Refer to A4 Inverter” on page 2-16.)
If any of the signals are incorrect:
a. Using a DVM, measure the signals shown in Table 1-5 at J30 of the A23 Motherboard.
If the signals in Table 1-5 are correct, replace the A3 Front Panel Auxilary Board.
(Refer to A3 Front Panel Auxilary Board” on page 2-14.)
If the 5.2 VD_LCD line is incorrect, refer to the “ILT-1 Power Supply Failure and Self-Test” on
page 1-4.
If either the LIGHT_EN or VDISP are incorrect:
a. Using a DVM, measure the signals at J15-37 (LIGHT_EN)
or J15-97 (VDISP) of the A23 Motherboard.
If either signal is incorrect, replace the A12 CPU.
(Refer to A12 CPU” on page 2-30.)
Table 1-4
Signal State
CN1-1 5.2 Vdc
CN1-3 LIGHT_EN >3 Vdc
CN1-4 VDISP Brightness = 50, VDSP < 50 mVdc
Brightness = 1, VDSP = 3 Vdc
Table 1-5
Signal State
J30-32 5.2 VD_LCD 5.2 Vdc
J30-20 LIGHT_EN >3 Vdc
J30-30 VDISP Brightness = 50, VDSP < 50 mVdc
Brightness = 1, VDSP = 3 Vdc