Installation guide
Assembly Replacement
A23BT1 Battery
2-55
A23BT1 Battery
Tools Required
•none
Removal Procedure
Refer to Figure 2-27 for this procedure.
1. Disconnect the power cord.
2. Remove the outer cover. Refer to “Outer and Inner Instrument Covers” on page 2-4.
3. Place the signal generator on its side so that the bottom of the signal generator is visible.
4. Remove the battery.
Replacement Procedure
1. Reverse the order of the removal procedure, but pay special attention when replacing the battery; angle
the left-side of the battery so that it goes under the positive prongs before pushing it down.
2. Perform the post-repair adjustments and performance tests that pertain to this removal procedure.
Figure 2-27 A23BT1 Battery
(Option 501, 502, 503, 504, Option 300, Option UNJ, or Option 506)