Installation guide

Assembly Replacement
A11 Internal Bit Error Rate Analyzer (Option UN7)
2-28
A11 Internal Bit Error Rate Analyzer (Option UN7)
Tools Required
needle-nose pliers
Removal Procedure
Refer to Figure 2-14 for this procedure and to the connector/cable diagram located on the card cage.
1. Disconnect the power cord.
2. Remove the outer cover. Refer to “Outer and Inner Instrument Covers” on page 2-4.
3. Using needle-nose pliers, disconnect the following cables:
W15 cable from A10J14
(This is W77 on an Option 1EM.)
W16 cable from A10J13
(This is W76 on an Option 1EM.)
W17 cable from A10J12
(This is W75 on an Option 1EM.)
W18 cable from A10J11
(This is W74 on an Option 1EM.)
W19 cable from A10J10
W20 cable from A10J9
W1 cable from A10J8
(This is W73 on an Option 1EM.)
W2 cable from A10J7
(This is W72 on an Option 1EM.)
W12 cable from A7J800.
(This is W78 on an Option 1EM.)
4. If you have an instrument with Option 506, use needle-nose pliers and disconnect the following cables:
W21 cable from A10J6
W22 cable from A10J5
5. Using needle-nose pliers, disconnect the following Option UN7 cables:
W23 from A11J4
W24 from A11J5
W25 from A11J6
6. Disconnect W10 from A7J13, A9J3, and A11J3.
7. Locate Top-SLOT 4, lift the retention levers, and disconnect the A11 Internal Bit Error Rate Analyzer
(Option UN7) from the A23 Motherboard.