Installation guide
Assembly Replacement
Before You Replace an Assembly
2-2
Before You Replace an Assembly
CAUTION Many of the assemblies in this instrument are very susceptible to damage from electrostatic
discharge (ESD). Perform service procedures only at a static-safe workstation and wear a
grounding strap.
Be sure to review the warning and caution statements described in Chapter 5, “Safety and Regulatory,” on
page 5-1 prior to replacing an assembly in your signal generator.
After Replacing an Assembly
After you have replaced or repaired an assembly, certain performance tests and adjustments may have to be
performed. Chapter 4, “Post-Repair Procedures,” on page 4-1, lists the performance tests and adjustments
required for each assembly.