Installation guide

Troubleshooting
Getting Started with Troubleshooting
1-4
ILT-1 Power Supply Failure and Self-Test
When the signal generator appears to be dead (no display or fan) there is often little evidence that points
directly to the cause. This section provides steps and solutions to typical failure modes.
CAUTION Immediately unplug the signal generator from the ac power line if the unit shows any of the
following symptoms:
Smoke, arcing, or unusual noise from inside the signal generator.
A circuit breaker or fuse on the main ac power line opens.
These potentially serious faults must be corrected before proceeding.
Power Supply LED Checks
Check the front panel LEDs (located next to the LINE switch) for normal operation in both standby and
power-on conditions. Normal operation is indicated as follows:
If normal operation is not observed, refer to “Power Supplies” on page 1-23.
If the fan does not operate when the signal generator is powered on, refer to the assembly level
troubleshooting section and verify the fan voltage.
If the fan operates when the signal generator is powered on,
refer to ILT-2 A12 CPU Power-On Test” on page 1-5.
Table 1-2
Power Switch Yellow LED Green LED
Standby On Off
Power-On Off On