Installation guide

Performance Tests
Phase Modulation Frequency Response
3-16
Phase Modulation Frequency Response
This automated test verifies the phase modulation frequency response specifications. The equipment
measures the variations in phase deviations due to changes in the applied ΦM rate; dc to 100 kHz. The
response is relative to a 1 kHz reference at the same phase deviation, and expressed in dB. Each frequency is
tested with the following sequence:
1. A reference signal is measured.
2. The set deviation is measured for each applied rate change.
3. Each deviation measurement is compared to the reference signal deviation.
The comparison is a calculation of the difference in the deviations relative to the reference rate. The relative
value is the ΦM frequency response.
Recommended Equipment
Agilent 8902A measuring receiver
Agilent 8904A function generator
Equipment Setup
Figure 3-14 Phase Modulation Frequency Response Setup