Installation guide

Performance Tests
AM Frequency Response
3-14
AM Frequency Response
This automated test verifies the AM frequency response specifications. The equipment measures the
variations in modulation depth due to changes in the applied AM rate; dc to 10 kHz. The response is relative
to a 1 kHz rate set at the test depth, and expressed in dB. Each frequency is tested with the following
sequence:
1. A reference signal is measured.
2. The set depth is measured for each applied rate change.
3. Each depth measurement is compared to the reference signal depth.
The comparison is a calculation of the difference in the depths relative to the reference rate. The relative
value is the AM Frequency Response.
Recommended Equipment
Agilent 8902A measuring receiver
Agilent 8904A function generator
Use one of the following:
for models with Option 501, 502, 503, 504 or Option UNJ, use an Agilent 8662A/8663A
high-performance RF signal generator
for models with Option 506, use an Agilent E4426/E4437 signal generator
MD/MDC-164 mixer (0.5 to 9 GHz)
Equipment Setups
Figure 3-12 AM Frequency Response 1300 MHz Setup