Installation guide

Adjustments
Power Level Accuracy, Low Power Calibration
4-37
Procedure
1. Connect the equipment as shown below.
2. Preset all of the equipment.
3. Follow the instructions as they appear on the controllers display.
Figure 4-27 Low-Power, Power Level Accuracy < 10 MHz Setup