Keysight Technologies E3640A - E3649A

dataTec ▪  Ferdinand-Lassalle-Str. 52  ▪  72770 Reutlingen  ▪  Tel. 07121 / 51 50 50  ▪  Fax 07121 / 51 50 10  ▪  info@datatec.de  ▪  www.datatec.de
4
Keysight E3640A – E3649A Programmable DC Power Supply Specications
Model number E3640A E3641A E3642A E3643A E3644A E3645A
Maximum power 30 W 50 W 80 W
# of outputs 1 1 1 1 1 1
DC output rating
(@ 0 ºC to 40 ºC)
0 to 8 V/3 A or
0 to 20 V/1.5 A
0 to 35 V/0.8 A or
0 to 60 V/0.5 A
0 to 8 V/5 A or
to 20 V/2.5 A
0 to 35 V/1.4 A or
0 to 60 V/0.8 A
0 to 8 V/8 A or
0 to 20 V/4 A
0 to 35 V/2.2 A or
0 to 60 V/1.3 A
Net weight 5.3 kg 5.2 kg 6.3 kg 6.2 kg 6.6 kg 6.7 kg
Dimensions (without bumper) 212.6 mm W x 88.5 mm H x 348.3 mm D
Model number E3646A E36 47A E3648A E3649A
Maximum power 60 W 100 W
# of outputs 2 2 2 2
DC output rating
(@ 0 ºC to 40 ºC)
Two 0 to 8 V/3 A or
0 to 20 V/1.5 A
Two 0 to 35 V/0.8 A or
0 to 60 V/0.5 A
Two 0 to 8 V/5 A or
0 to 20 V/2.5 A
Two 0 to 35 V/1.4 A or
0 to 60 V/0.8 A
Net weight 8.2 kg 8.0 kg 9.2 kg 9.1 kg
Dimensions (without bumper) 212.8 mm W x 133.0 mm H x 348.3 mm D
Load and line regulation ± (% of output + offset)
Voltage < 0.01% + 3 mV
Current < 0.01% + 250 μA
Ripple and noise (20 Hz to 20 MHz)
Normal mode voltage < 5 mVpp/0.5 mVrms for 8 V/20 V models
< 8 mVpp/1 mVrms for 35 V/60 V models
Normal mode current < 4 mArms
Common mode current < 1.5 μArms
Accuracy
1
12 months (@ 25 ºC ± 5 ºC), ± (% output + offset)
Programming
Voltage < 0.05% + 10 mV (< 0.1% + 25 mV for output 2 of E3646/47/48/49 A)
Current < 0.2% + 10 mA
Readback (over GPIB and with respect to actual output)
Voltage < 0.05% + 5 mV (< 0.1% + 25 mV for output 2 of E3646/47/48/49 A)
Current < 0.15% + 5 mA (< 0.15% + 10 mA for output 2 of E3646/47/48/49 A)
Meter
2
(over front panel with respect to actual output)
Voltage < 0.05% + 2 counts (<0.1% + 4 counts for output 2)
Current < 0.15% + 5 mA (<0.15% + 10 mA for output 2)
Resolution
Program < 5 mV/1 mA
Readback < 2 mV/1 mA
Meter 10 mV/1 mA
Transient response Less than 50 μsec for output to recover to within 15 mV following a change in output current from full
load to half load or vice versa.
Settling time
3
< 90 msec
OVP
Accuracy,
± (% output + offset)
Activation time
4
< 0.5% + 0.5 V
< 1.5 msec, OVP ¥3 V/< 10 msec, OVP < 3 V
1. Accuracy specications are valid after a 1-hour warm-up with no load and calibration at 25 ºC.
2. Meter accuracy specication is at minimum 10 mV decimal limited by front panel resolution.
3. Maximum time required for the output voltage to change from 1% to 99% or vice versa following the receipt of VOLTage or APPLy command via direct GPIB
or RS-232 interface.
4. Average time for output to start and drop after an OVP condition occurred.
04 | Keysight | E3640A – E3649A Programmable DC Power Supplies – Data Sheet