Technical data

2 Operation and Features
Programming the Overvoltage Protection
44 E364xA User’s and Service Guide
By adjusting the output voltage level
i Lower the output voltage level below the OVP trip point.
The OVP and Limit annunciators are flashing after
is pressed.
ii Check that you lowered the voltage level below the OVP
trip point.
The OVP trip point is displayed. Do not adjust the trip
point at this step.
iii Select the OVP CLEAR mode by turning the knob.
OVP ON
OVP CLEAR
iv Clear the overvoltage condition and exit this menu.
DONE
The OVP annunciator will not flash any more. The
output will return to the meter mode.
By adjusting the OVP trip level
i Raise the OVP trip level higher than the level tripped.
ii Select the OVP CLEAR mode by turning the knob.
OVP ON
OVP CLEAR
iii Clear the overvoltage condition and exit this menu.
DONE
The OVP annunciator will not flash any more. The
output will return to the meter mode.
Remote interface operation
Display
Limit
Display
Limit
Over
Voltage
Over
Voltage
Over
Voltage
Over
Voltage
Over
Voltage
Over
Voltage
VOLT:PROT {<voltage>|MIN|MAX} Set the OVP level.