Technical data

8 Service and Maintenance
General Calibration or Adjustment Procedure
210 E364xA User’s and Service Guide
12 Enter the computed value (DVM reading divided by the
shunt resistance) by using the knob and resolution
selection keys. Notice that you should wait for the DVM
reading to stabilize for accurate calibration during the
current calibration.
I LO 0.1900
13 Save the changes and select the middle current calibration
point.
I MI 1.5000
If the entered number is within an acceptable range, an
ENTERED message appears for a second. If the entered
number is not correct, an error message will be displayed
for a second and you will hear a beep, and then return to
the low, middle, or high current calibration point again.
14 Enter the computed value (DVM reading divided by the
shunt resistance) by using the knob and resolution
selection keys.
I MI 1.5400
15 Save the change and select the high current calibration
point.
I HI 2.8000
16 Enter the computed value (DVM reading divided by the
shunt resistance) by using the knob and resolution
selection keys.
I HI 2.7890 A
17 Save the new current calibration constants and exit the
calibration mode.
CAL MODE
If the calibration fails, an error message will be displayed
for one second and you will hear a beep, and then return
to the current calibration mode again. A CURRENT CAL
message is displayed.
Calibrate
View
Calibrate
View
Calibrate
View
Power