Technical data
Service and Maintenance 8
General Calibration or Adjustment Procedure
E364xA User’s and Service Guide 209
6 Enter the reading you obtained from the DVM by using
the knob and resolution selection keys.
V MI 11.058
7 Save the changes and select the high voltage calibration
point.
V HI 19.500
8 Enter the reading you obtained from the DVM by using
the knob and resolution selection keys.
V HI 19.495
9 Save the changes and go to the OVP calibration mode.
OVP CAL
If the calibration fails, an error message will be displayed
for a second and return to the voltage calibration mode
again. A VOLTAGE CAL message is displayed. The message
above is displayed to indicate that the power supply is
ready for the OVP calibration.
10 Run the OVP calibration.
CALibrating
The above message is displayed to indicate that the
calibration is in progress. It takes approximately
10 seconds to complete the calibration. If the calibration
fails, an error message will be displayed for a second and
you will hear a beep, and then return to the OVP
calibration mode again.
Current calibration
11 Select the low- end current calibration point.
CURRENT CAL
I LO 0.2000
Calibrate
View
Calibrate
View
Calibrate
View
NOTE
Connect an appropriate shunt 0.01 Ω across the output terminals, and
connect a digital voltmeter across the shunt resistor for the current
calibration.
Calibrate
View