Technical data
Service and Maintenance 8
Constant Current (CC) Verifications
E364xA User’s and Service Guide 199
CC PARD (ripple and noise)
Periodic and random deviations (PARD) in the output (ripple
and noise) combine to produce a residual AC current, as
well as an AC voltage superimposed on the DC output. CC
PARD is specified as the rms output current in a frequency
range of 20 Hz to 20 MHz with the power supply in the
constant current operation.
E3640A E3641A E3642A E3643A E3644A E3645A
0.55 mA 0.33 mA 0.75 mA 0.39 mA 1.05 mA 0.47 mA
NOTE
For better measurement results, it is recommended to make the
connection between the BNC receptacle and the output terminals as short
as possible, and to use the recommended split ferrites with the
BNC-to-BNC cable as shown in Figure 8-5.
Input
Rms voltmeter
Rms voltmeter
(Front Panel Connections)
(Rear Panel Connections)
Current
monitoring
resistor
Current
monitoring
resistor
BNC cable
BNC
receptacle
Split
ferrites
Input