Technical data

Service and Maintenance 8
Constant Voltage (CV) Verifications
E364xA User’s and Service Guide 187
Constant Voltage (CV) Verifications
Constant voltage test setup
If more than one meter or if a meter and an oscilloscope are
used, connect each to the (+) and (–) terminals by a
separate pair of leads to avoid mutual coupling effects. Use a
coaxial cable or a shielded 2- wire cable to avoid noise
pick- up on the test leads.
Voltage programming and readback accuracy
This test verifies that the voltage programming and GPIB or
RS- 232 readback functions are within specifications.
Table 8 - 5 Verification programming values
Model Low voltage range High voltage range
E3640A 8 V/3 A 20 V/1.5 A
E3641A 35 V/0.8 A 60 V/0.5 A
E3642A 8 V/5 A 20 V/2.5 A
E3643A 35 V/1.4 A 60 V/0.8 A
E3644A 8 V/8 A 20 V/4 A
E3645A 35 V/2.2 A 60 V/1.3 A
NOTE
The readback values over the remote interface should be identical to
those displayed on the front panel.
You should program the power supply over the remote interface for this
test to avoid round off errors.