Technical data
7 Characteristics and Specifications
Supplemental Characteristics
166 E364xA User’s and Service Guide
Supplemental Characteristics
Table 7 - 3 Supplemental characteristics
Parameter E3640A E3641A E3642A E3643A E3644A E3645A
Output
programming
range
(maximum
programmable
values)
Low range 0 to +8.24 V
0 to 3.09 A
0 to +36.05 V
0 to 0.824 A
0 to +8.24 V
0 to 5.15 A
0 to +36.05 V
0 to 1.442 A
0 to +8.24 V
0 to 8.24 A
0 to +36.05 V
0 to 2.266 A
High range 0 to +20.6 V
0 to 1.545 A
0 to +61.8 V
0 to 0.515 A
0 to +20.6 V
0 to 2.575 A
0 to +61.8 V
0 to 0.824 A
0 to +20.6 V
0 to 4.12 A
0 to +61.8 V
0 to 1.339 A
OVP 1 V to 22 V 1 V to 66 V 1 V to 22 V 1 V to 66 V 1 V to 22 V 1 V to 66 V
Remote sensing
capability
Voltage drop Up to 1 V per each lead
Load
regulation
Add 5 mV to specification for each 1 V change in the + output lead due to load current
changes.
Load voltage Subtract the voltage drop in load leads from the specified output voltage rating.
Temperature
coefficient
[1]
±(% of output
+ offset)
Voltage <0.01% + 3 mV
Current <0.02% + 3 mA
Stability
[2]
±(% of output +
offset)
Voltage <0.02% + 2 mV
Current <0.1% + 1 mA
Output voltage
overshoot
During turn-on or turn-off of AC power, output plus overshoot will not exceed 1 V if the
output control is set to less than 1 V. If the output control is set to 1 V or higher, there is no
overshoot.
Programming
language
Standard Commands for Programmable Instruments (SCPI)
State storage
memory
Five user-configurable stored states
Recommended
calibration
interval
1 year