Technical data
Characteristics and Specifications 7
Electrical Specifications
E364xA User’s and Service Guide 165
Load regulation
±(% of output + offset)
Voltage <0.01% + 3 mV
Current <0.01% + 250 μA
Line regulation
±(% of output + offset)
Voltage <0.01% + 3 mV
Current <0.01% + 250 μA
Programming
resolution
Voltage <5 mV
Current <1 mA
Readback resolution Voltage <2 mV
Current <1 mA
Front panel
resolution
Voltage 10 mV
Current 1 mA
Transient response
time
<50 μs for the output to recover to within 15 mV following a change in the output
current from full load to half load or vice versa.
Settling time <90 ms for the output voltage to change from 1% to 99% or vice versa following the
receipt of the VOLTage or APPLy command via direct GPIB or RS-232 interface.
OVP accuracy
±(% of output + offset)
<0.5% + 0.5 V
Activation time
(Average time for the
output to start to drop
after the OVP condition
occurs)
<1.5 msec when the trip voltage is equal or greater than 3 V
<10 msec when the trip voltage is less than 3 V
[1] Accuracy specifications after a 1-hour warm-up period with no load and calibration at 25 °C.
[2] Meter accuracy specification at a minimum of 10 mV decimal limited by the front panel resolution.
Table 7 - 2 Electrical specifications (continued)
Parameter E3640A E3641A E3642A E3643A E3644A E3645A