Technical data
6 Tutorial
Output Characteristics
154 E364xA User’s and Service Guide
Figure 6-6 Simplified diagram of the common mode and normal mode
sources of noise
When the load changes very rapidly, as when a relay contact
is closed, the inductance in the hook up wire and in the
power supply output will cause a spike to appear at the
load. The spike is a function of the rate of change of the
load current. When very rapid changes in load are expected,
a capacitor with a low series resistance, in parallel with the
power supply, and close to the load is the best way to
minimize these voltage spikes.
OUTPUT
TERMINAL
OUTPUT
VOLTAGE
COMMON
MODE
NOISE
<1.5 μArms
NORMAL
MODE
NOISE
R
<5 mVpp*
<0.5 mVrms*
<8 mVpp**
<1 mVrms**
*E3640A/E3642A/E3644A models
**E3641A/E3643A/E3645A models