Technical data
Error Messages 4
Execution Error Messages
E364xA User’s and Service Guide 127
-113 Undefined header
A command was received that is not valid for this power supply. You may have misspelled the command
or it may not be a valid command. If you are using the short form of the command, remember that it may
contain up to four letters.
Example: TRIGG:DEL 3
-114
[1]
Header suffix out of range
The numeric suffix attached to a command header is not one of the allowable values.
Example: OUTPUT2 ON
-120
[1]
Numeric data error
An invalid number was specified for a numeric parameter.
Example: APPL 1.0E+320000
-121 Invalid character in number
An invalid character was found in the number specified for a parameter value.
Example: *ESE #B01010102
-123 Numeric overflow
A numeric parameter was found whose exponent was larger than 32000.
-124 Too many digits
A numeric parameter was found whose mantissa contained more than 255 digits, excluding leading
zeros.
-128 Numeric data not allowed
A numeric parameter was received but a character string was expected.
Example: DISP:TEXT 123
-130
[1]
Suffix error
A suffix was incorrectly specified for a numeric parameter. You may have misspelled the suffix or the
numeric parameter does not accept a suffix.
Example: TRIG:DEL 0.5 SECS
-131 Invalid suffix
A suffix was incorrectly specified for a numeric parameter. You may have misspelled the suffix.
Example: TRIG:DEL 0.5 SECS
-134 Suffix too long
A suffix for a numeric parameter contained too many characters.
Table 4 - 1 Execution error messages (continued)