Technical data

4 Error Messages
Overview
124 E364xA User’s and Service Guide
Overview
Errors are retrieved in the first- in- first- out (FIFO) order.
The first error returned is the first error that was stored.
Errors are cleared as you read them. When you have read all
errors from the queue, the ERROR annunciator turns off and
the errors are cleared. The power supply beeps once each
time an error is generated.
If more than 20 errors have occurred, the last error stored
in the queue (the most recent error) is replaced with
- 350, “Queue overflow”. No additional errors are stored until
you remove errors from the queue. If no errors have
occurred when you read the error queue, the power supply
responds with +0, “No error” over the remote interface or NO
ERRORS from the front panel.
The error queue is cleared by the *CLS (clear status)
command or when power is cycled. The errors are also
cleared when you read the queue.
Front panel operation
Press the (Local) key to return to the front panel
operation mode if you operate the power supply in the
remote operation mode.
ERRORS
If the ERROR annunciator is turned on, press to view
the errors. Use the knob to scroll through the error
numbers. Press to view the text of the error message.
Press to increase the scrolling speed of the text on the
display. All errors are cleared when you exit the menu by
pressing or let the display time- out for approximately
30 seconds.
NOTE
The *RST (reset) command does not clear the error queue.
Local
Store
View
View
>
>
View