Technical data
Remote Interface Reference 3
Calibration Commands
E364xA User’s and Service Guide 101
14 Record calibration information such as the next
calibration due date or contact person for future
reference. The calibration string may contain up to 40
characters.
CALibration:STRing ‘<string>’
NOTE
You should wait for the DVM reading to stabilize for accurate calibration.