Technical data
Remote Interface Reference 3
Output Settings and Operation Commands
E364xA User’s and Service Guide 87
Before sending this command, lower the output voltage
below the OVP trip point, or raise the OVP trip level above
the output setting. Note that the overvoltage condition
caused by an external source must be removed first before
proceeding with this command.
VOLTage:RANGe
{P8V
[1]
|P20V
[1]
|P35V
[2]
|P60V
[2]
|LOW|HIGH}
Select an output range to be programmed by the identifier.
For example, P20V or HIGH is the identifier for the
20 V/1.5 A range and P8V or LOW is for the 8 V/3 A range
(for the E3640A model). At *RST, the low voltage range is
selected.
VOLTage:RANGe?
Query the presently selected range. The returned parameter
is P8V or P35V for the low voltage range, or P20V or P60V
for the high voltage range.
MEASure:CURRent?
Query the current measured across the current sense
resistor inside the power supply.
MEASure[:VOLTage]?
Query the voltage measured at the sense terminals of the
power supply.
[1] For the E3640A/E3642A/E3644A models.
[2] For the E3641A/E3643A/E3645A models.