Technical data
Remote Interface Reference 3
Simplified Programming Overview
E364xA User’s and Service Guide 79
Power supply programming ranges
The SOURce subsystem requires parameters for programming
values. The available programming value for a parameter
varies according to the desired output range of the power
supply. Table 3- 2 lists the programming values available and
the MINimum, MAXimum, DEFault, and reset values of your
power supply.
Refer to Table 3- 2 and Table 3- 3 to identify the
programming values when programming the power supply.
Table 3 - 2 Agilent E3640A/E3642A/E3644A programming ranges
E3640A E3642A E3644A
0 to 8 V/3 A
range
0 to 20 V/1.5 A
range
0 to 8 V/5 A
range
0 to 20 V/2.5 A
range
0 to 8 V/8 A
range
0 to 20 V/4 A
range
Voltage Programming
range
0 V to 8.24 V 0 V to 20.60 V 0 V to 8.24 V 0 V to 20.60 V 0 V to 8.24 V 0 V to 20.60 V
MAX value 8.24 V 20.60 V 8.24 V 20.60 V 8.24 V 20.60 V
MIN value 0 V 0 V 0 V
DEFault value 0 V 0 V 0 V
*RST value 0 V 0 V 0 V
Current Programming
range
0 A to 3.09 A 0 A to 1.545 A 0 A to 5.15 A 0 A to 2.575 A 0 A to 8.24 A 0 A to 4.12 A
MAX value 3.09 A 1.545 A 5.15 A 2.575 A 8.24 A 4.12 A
MIN value 0 A 0 A 0 A
DEFault value 3 A 1.5 A 5 A 2.5 A 8 A 4 A
*RST value 3.00 A 5.00 A 8.00 A