User`s guide
Table Of Contents
- Agilent E3632A DC Power Supply
- Table of Contents
- List of Figures
- List of Tables
- 1 Getting Started
- 2 Operation and Features
- Overview
- Constant Voltage Operation
- Constant Current Operation
- Storing and Recalling Operating States
- Programming the Overvoltage Protection
- Programming the Overcurrent Protection
- Remote Voltage Sensing
- Disabling the Output
- System-Related Operations
- Remote Interface Configuration
- GPIB Interface Configuration
- RS-232 Interface Configuration
- Calibration
- 3 Remote Interface Reference
- SCPI Command Summary
- Introduction to the SCPI Language
- Simplified Programming Overview
- Using the APPLy Command
- Output Settings and Operation Commands
- Triggering Commands
- System-Related Commands
- Calibration Commands
- RS-232 Interface Commands
- SCPI Status Registers
- What is an event register?
- What is an enable register?
- SCPI status system
- The Questionable Status register
- The Standard Event register
- The Status Byte register
- Using Service Request (SRQ) and Serial POLL
- Using *STB? to read the Status Byte
- Using the Message Available Bit (MAV)
- To interrupt your bus controller using SRQ
- To determine when a command sequence is completed
- Using *OPC to signal when data is in the output buffer
- Status Reporting Commands
- Halting an Output in Progress
- SCPI Conformance Information
- IEEE-488 Conformance Information
- 4 Error Messages
- 5 Application Programs
- 6 Tutorial
- 7 Characteristics and Specifications

3 Remote Interface Reference
System-Related Commands
96 E3632A User’s Guide
System-Related Commands
DISPlay {OFF|ON}
Turn the front- panel display off or on. When the display is
turned off, outputs are not sent to the display and all
annunciators are disabled except the ERROR annunciator.
The display state is automatically turned on when you
return to the local mode. Press to return to the local
state from the remote interface.
DISPlay?
This query returns the front- panel display setting. Returns 0
(OFF) or 1 (ON).
DISPlay:TEXT <quoted string>
Display a message on the front panel. The power supply will
display up to 12 characters in a message; any additional
characters are truncated. Commas, periods, and semicolons
share a display space with the preceding character, and are
not considered individual characters.
DISPlay:TEXT?
This query returns the message sent to the front panel and
returns a quoted string.
DISPlay:TEXT:CLEar
Clear the message displayed on the front panel.
OUTPut {OFF|ON}
Enable or disable the outputs of the power supply. When the
output is disabled, the voltage value is 0 V and the current
value is 20 mA. At *RST, the output state is OFF.
Local