Technical data

Service 2
To Disconnect the Output Using an External Relay
E3632A Service Guide 53
Installation procedure for an external relay
The assembly drawings are located in “Component Locator
Diagram” on page 61.
1 Remove the front and rear bumpers and take off the
cover.
2 Install JP3
[1]
and JP4
[1]
located adjacent to the connector
P5 (refer to “Component locator diagram for the main
board assembly — top” on page 61). A bare wire may be
used.
3 Reassemble the power supply.
[1] For serial MY53xx6xxx, install JP751 and JP752 located adjacent to the
connector CN751 (refer to “Component locator diagram for the main board
assembly — top (serial MY53xx6xxx)” on page 63)