Technical data
Calibration Procedures 1
General Calibration/Adjustment Procedure
E3632A Service Guide 35
19 Pressing the Calibrate key saves the new calibration
constants for the output current and goes to the OCP
calibration mode.
CAL SETUP 4
A CALIBRATING message appears for one second to
indicate that the current calibration is progressing and
new calibration constants of “SETUP 3” are stored. Then,
the display shows the above message to indicate that the
power supply is ready for the OCP calibration.
If the calibration fails, a DAC CAL FAIL or ADC CAL FAIL
message appears for one second and the display shows
the CAL SETUP 3 for current calibration again.
20 Pressing the Calibrate key saves the new OCP calibration
constants and return to the calibration mode.
CAL MODE
A CALIBRATING message appears for several seconds to
indicate that the OCP calibration is progressing and new
OCP calibration constants of “SETUP 4” are stored. Then
the display will return to the calibration mode.
21 Turn off the power supply to exit the calibration mode.
NOTE
Notice that you should wait for the DVM reading to stabilize for accurate
calibration.
Calibrate
Calibrate
Power