Technical data
1 Calibration Procedures
General Calibration/Adjustment Procedure
34 E3632A Service Guide
15 Pressing the Calibrate key saves the change and selects the
middle current calibration point.
I MI 3.5000 A
If the entered number is within an acceptable range, an
ENTERED message appears for one second. If the entered
number is not correct, a MIN VALUE or MAX VALUE message
appears for one second and the display shows the low
current calibration point again. The display now shows
the middle current calibration point.
16 Read the DVM and change the middle current value on the
display to match the computed current (DVM reading
divided by shunt resistance).
For example, if the computed value is 3.499 A, adjust the
current to 3.499 A using the knob and arrow keys.
I MI 3.4990 A
17 Pressing the Calibrate key saves the change and selects the
high current calibration point.
I HI 6.9000 A
If the entered number is within an acceptable range, an
ENTERED message appears for one second. If the entered
number is not correct, a MIN VALUE or MAX VALUE message
appears for one second and the display shows the middle
current calibration point again. The display now shows
the high current calibration point.
18 Read the DVM and change the high current value on the
display to match the computed current (DVM reading
divided by shunt resistance).
For example, if the computed value is 6.899 A, adjust the
current to 6.899 A using the knob and arrow keys.
I HI 6.8990 A
Calibrate
NOTE
Notice that you should wait for the DVM reading to stabilize for accurate
calibration.
Calibrate