Technical data
Calibration Procedures 1
General Calibration/Adjustment Procedure
E3632A Service Guide 31
5 Calibrate DAC and select the low voltage calibration point.
30 LEFT
The START BITCAL message is displayed for about
3 seconds to indicate that the power supply is ready for
DAC calibration. Then it counts down numbers from
30 to 0.
V LO 0.5000 V
Then, the display shows the low voltage calibration point.
6 Read the DVM and change the low voltage value on the
display to match the measured voltage.
For example, if the DVM reading is 0.4500 V, adjust the
voltage to 0.4500 V using the knob and resolution
selection keys.
V LO 0.4500 V
7 Pressing the Calibrate key saves the change and selects the
middle voltage calibration point.
V MI 15.000 V
If the entered number is within an acceptable range, an
ENTERED message appears for one second. If the entered
number is not correct, a MIN VALUE or MAX VALUE message
appears for one second and the display shows the low
voltage calibration point again. The display now shows the
middle voltage calibration point.
8 Read the DVM and change the middle voltage value on the
display to match the measured voltage.
For example, if the DVM reads 14.995 V, adjust the voltage
to 14.995 V using the knob and arrow keys.
V MI 14.995 V
Calibrate
Calibrate