Technical data
Calibration Procedures 1
General Calibration/Adjustment Procedure
E3632A Service Guide 29
General Calibration/Adjustment Procedure
The calibration procedures from the front panel are
described in this section. For voltage calibration, disconnect
all loads from the power supply and connect a DVM across
the output terminals. For current calibration, disconnect all
loads from the power supply, connect an appropriate current
monitoring resistor (0.01 Ω) across the output terminals, and
connect a DVM across the terminals of the monitoring
resistor.
The following table shows calibration parameters and points
which should be used to calibrate the output voltage and
current.
NOTE
The power supply should be calibrated after 1-hour warm-up with no load
connected.
Table 1 - 4 Parameters for calibration
Calibration parameter Voltage/current Calibration point mnemonic
CAL SETUP 1 Voltage V LO
V MI
V HI
CAL SETUP 2 OVP None
CAL SETUP 3 Current I LO
I MI
I HI
CAL SETUP 4 OCP None