Technical data
1 Calibration Procedures
Calibration Security Code
26 E3632A Service Guide
4 Unsecure the power supply.
UNSECURED
The power supply is unsecured when you press the
key. You will see the above message from the
front panel for one second. The CAL MODE message is
displayed on the front panel after above message.
5 Turn off the calibration mode.
Turn off the power supply to exit the calibration mode.
To unsecure the power supply without the security code
To unsecure the power supply without the correct security
code (when you forget the security code), follow the steps
below. Refer to “Electrostatic Discharge (ESD)
Precautions” on page 51 before beginning this procedure.
1 Disconnect the power cord and all load connections from
front terminals.
2 Remove the instrument cover.
3 Connect the power cord and turn on the calibration mode
by pressing the Calibrate key while simultaneously turning
on the power supply, and then continue to hold the
Calibrate key for about 5 seconds until a beep is heard. Be
careful not to touch the power line connections.
4 Apply a short between the two exposed metal pads on
JP5
[1]
(located near U13). Refer to “Component locator
diagram for the main board assembly — top” on page 61.
5 While maintaining the short, press to move to the
security code and enter any unsecure code in the
calibration mode. The power supply is now unsecured.
Secure
Secure
Power
NOTE
To re-secure the power supply (following calibration), perform this
procedure again.
[1] For serial MY53xx6xxx, apply the short at JP12 (located near U36). Refer to
“Component locator diagram for the main board assembly — top (serial
MY53xx6xxx)” on page 63.
Secure