Technical data

Calibration Procedures 1
Performance Test Record for E3632A
E3632A Service Guide 23
CC performance test record
Table 1 - 3 CC performance test record
Test Description Actual Result Specifications
Upper Limit Lower Limit
CC programming accuracy @ 0 A (I
O
) +0.0100 A –0.0100 A
CC readback accuracy @ 0 A I
O
+ 0.0050 A I
O
– 0.0050 A
CC programming accuracy @ full scale (I
O
) 7.0240 A 6.9760 A
CC readback accuracy @ full scale I
O
+ 0.0155 A I
O
– 0.0155 A
CC load regulation Maximum change: <0.95 mA
CC line regulation Maximum change: <0.95 mA
CC ripple/noise <2 mA
Common mode current noise <1.5 μArms