Technical data
1 Calibration Procedures
Constant Current (CC) Verifications
20 E3632A Service Guide
3 The output current should be at the full- rated rating with
the CC annunciator on. If not lit, adjust the load so that
the output voltage drops slightly until the CC annunciator
lights.
4 Divide the reading on the RMS voltmeter by the load
resistance to obtain rms current. The readings should be
within the limit of 2 mA.