[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]
{<current>| MINimum | MAXimum}
This command programs the pending triggered current level of the power
supply. The pending triggered current level is a stored value that is
transferred to the output terminals when a trigger occurs. A pending
triggered level is not affected by subsequent
CURRent
commands.
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]?
[MINimum | MAXimum]
This query returns the presently programmed triggered current level. If no
triggered level is programmed, the
CURRent
level is returned.
CURRent
:TRIGgered? MAXimum
and
CURRent:TRIGgered? MINimum
return the
maximum and minimum programmable triggered current levels.
VOLTage[:LEVel][:IMMediate][:AMPLitude]
{<voltage>| MINimum | MAXimum}
This command directly programs the immediate voltage level of the power
supply. The immediate level is the voltage limit value of the selected output
with the
INSTrument
command.
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]?
[MINimum | MAXimum]
This query returns the presently programmed voltage limit level of the
selected output.
VOLTage? MAXimum
and
VOLTage? MINimum
return
the maximum and minimum programmable voltage levels of the selected
output.
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]
{<voltage>| MINimum | MAXimum}
This command programs the pending triggered voltage level of the power
supply. The pending triggered voltage level is a stored value that is
transferred to the output terminals when a trigger occurs. A pending
triggered level is not affected by subsequent
VOLTage
commands.
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]?
[MINimum | MAXimum]
This query returns the presently programmed triggered voltage level. If no
triggered level is programmed, the
VOLTage
level is returned.
VOLTage
:TRIGgered? MAXimum
and
VOLTage:TRIGgered? MINimum
return
the maximum and minimum programmable triggered voltage levels.