Specifications
74 Chapter 4
Preparation for Accurate Measurement
Calibration
Calibration
To assure accurate measurements, calibration must be performed on the DUT port on the
test head. In the Dielectric Material measurement mode, calibration should be performed
on the surface of the test fixture that makes contact with the DUT. The following section
explains how to perform calibration.
Define Calibration Kit
Selection
Use the following GPIB command to select whether the 7-mm calibration kit included with
the E4991A or a user-defined calibration kit is used for calibration.
• SENS:CORR1:CKIT on page 420
How to Enter Standard Values for User-Defined Calibration Kit
When a user-defined calibration kit is used to perform calibration, use the following GPIB
command to select whether the Open/Short/Load standard values are set to a single value
or different values for each frequency.
• SENS:CORR1:CKIT:LIST on page 421
Standard Value for User-Defined Calibration Kit (Single Value)
Use the following GPIB commands to define the Open/Short/Load standard values as a
single value common to each frequency.
Standard Value Array for User-Defined Calibration Kit (Frequency)
By defining the Open/Short/Load standard values of the calibration kit as an array for each
frequency, errors due to frequency characteristics of the Open/Short/Load standard can be
removed. Use the following GPIB commands to define standard values as an array for each
frequency.
Standard Parameter GPIB command
Open Conductance value (G) SENS:CORR1:CKIT:STAN1:G on page 423
Capacitance value (C) SENS:CORR1:CKIT:STAN1:C on page 422
Short Resistance value (R) SENS:CORR1:CKIT:STAN2:R on page 429
Inductance value (L) SENS:CORR1:CKIT:STAN2:L on page 426
Load Resistance value (R) SENS:CORR1:CKIT:STAN3:R on page 433
Inductance value (L) SENS:CORR1:CKIT:STAN3:L on page 430
Standard Parameter GPIB command
Open Conductance value (G) SENS:CORR1:CKIT:STAN1:LIST:G on page 425
Susceptance value (B) SENS:CORR1:CKIT:STAN1:LIST:B on page 424
Short Resistance value (R) SENS:CORR1:CKIT:STAN2:LIST:R on page 427
Reactance value (X) SENS:CORR1:CKIT:STAN2:LIST:X on page 428
Load Resistance value (R) SENS:CORR1:CKIT:STAN3:LIST:R on page 431
Reactance value (X) SENS:CORR1:CKIT:STAN3:LIST:X on page 432