Specifications
64 Chapter 3
Setting Measurement Conditions
Combination of Several Sweep Conditions (Segment Sweep)
Combination of Several Sweep Conditions (Segment Sweep)
The Agilent E4991A allows you to set combinations of up to 16 segments, each with its
own sweep conditions, to perform segment sweep. A segment sweep table must be created
before segment sweep can be performed.
Create/Edit Segment Sweep Table
The segment sweep table can be created or edited in two ways: specifying each segment
parameter individually or specifying parameters collectively.
Display Segment Sweep Table
Use the following GPIB command to display the sweep conditions on the segment sweep
table.
• DISP:TEXT{1-3}:SET on page 356
Create Segments and Read Total Number of Segments
To create a segment, use the following GPIB command to create the specified number of
segments in the segment sweep table to make them available for editing. This command
clears the existing segments and creates new segments. When this command is used in the
query format, the total number of segments contained in the table is read.
• SEGM:COUN on page 397
Delete All Segments
Use the following GPIB command to delete all of the segments in the segment sweep table.
• SEGM:DEL:ALL on page 406
Select Oscillator Level Unit
Use the one of the following GPIB commands to select the unit (current/voltage/power) in
which the oscillator level is set when the segment sweep table is created.
Select DC Bias Output Mode (DC Bias Function)
When the dc bias function is used, use one of the following GPIB commands to select
whether the dc bias output level is set to the fixed voltage or fixed current source mode
when the segment sweep table is created.
Oscillator level GPIB command
Set in current SEGM:CURR:STAT on page 402
Set in voltage SEGM:VOLT:STAT on page 418
Set in power (dBm) SEGM:POW:STAT on page 412
Output mode GPIB command
Fixed current source mode SEGM:CURR:OFFS:STAT on page 399
Fixed voltage source mode SEGM:VOLT:OFFS:STAT on page 417