Specifications
Appendix G 611
Messages
Error number: 62
G. Messages
62 Can’t execute data examination
Even though the data for statistic analysis has been acquired, an attempt has been made to
read out the statistic analysis results by changing the setting conditions (such as sweep start
rate) and using
CALC{1-5}:MST:DATA? command on page 344 before the measurement
have been updated.
After changing the setting conditions, wait for the measurement to be updated, and then
read out the statistic analysis results.
69 Too many segments or points
During editing the list sweep table, setting has been attempted which will exceed the
maximum figure of segment (16), maximum figure of measurement points per 1 segment,
or maximum figure of total measurement points in all segments.
When setting, number of segments or measurement points should not exceed the maximum
figures.
70 Not allowed in this measurement mode
At present a command has been sent which is not possible to be executed in the set
measurement mode. For example, in case of Dielectric Measurement Mode,
SENS:CORR1:CKIT command on page 420 has been used in order to set the calibration
kit to used defined calibration kit.
Select a measurement mode will validate the command.
71 Impedance measurement mode only
A command has been sent which is valid for Impedance Measurement Mode only. For
example, during the Magnetic Substance Measurement Mode,
SENS:CORR2:CKIT:STAN1:C command on page 441 has been used in order to define
the user defined fixture compensation kit.
Select Impedance Measurement Mode.
72 Permittivity measurement mode only
A command has been sent which is valid for Dielectric Measurement Mode only. For
example, during the Magnetic Substance Measurement Mode,
SENS:CORR1:CKIT:STAN7:THIC command on page 436 has been used in order to set
the thickness of load standard for Dielectric Measurement.
Select Dielectric Measurement Mode.
73 Permeability measurement mode only
A command has been sent which is valid for Magnetic Substance Measurement Mode only.
For example, during the Dielectric Measurement Mode,
CALC:FORM:PAR:MAG
command on page 302 has been used in order to set the size of the magnetic substance.
Select Magnetic Substance Measurement Mode.
77 Invalid material size
The definition of the size of test material in the Magnetic Substance Measurement is
invalid. For example, outside diameter of the test material was attempted to set in smaller
figure than the inside diameter.
Set the outside diameter of the magnetic substance bigger than that of the inside diameter.